The P90 uses a flashing display to indicate specific functional failures. Error Code Likely Fix T-value is lower than actual temperature Adjust the program's holding temperature. Er03 Vacuum switch-off value exceeds temperature Ensure vacuum settings do not exceed preset temperature T. Er06 B-value (standby) is out of range Enter a standby temperature between 100°C and 700°C. Er12 Low battery in control unit Replace the internal battery immediately. Er13 Excessive temperature detected Check for programming errors or cooling system blockages. Operational Quick Guide
Use the green START key to begin a program and the red STOP key to abort it or silence the buzzer. ivoclar programat p90 manual fixed
Turn the furnace on and wait for the buzzer to sound before placing ceramics on the firing mount. The P90 uses a flashing display to indicate
Ensuring accurate firing temperature is critical. Use the + and = calibration keys along with the ATK2 automatic double-range calibration set to verify accuracy. Er06 B-value (standby) is out of range Enter